Specifications
14. Frequency
Span Accuracy
Specification
Related Performance
Test
Description
For spans
>l
MHz;
f(2%
of the actual frequency separation, + 0.5%
of span setting)
For span
51
MHz;
f(5%
of the actual frequency separation, + 0.5% of
span setting)
Frequency Span Accuracy Test
A stable signal source is connected to the spectrum analyzer and the
center frequency and span are set to measure spans of 10 MHz, 1
MHz, and 100 kHz.
5-24 Operation Verification