11. Sweep Time Accuracy
11. Sweep Time
Accuracy
Specification
Sweep time
1100
s
...............................
4110%
Sweep time
>lOO
s
...............................
f20%
Related Performance
Sweep Time Accuracy Test
Test
Description
An internal measurement function is used to measure the sweep
time at 20 ms, 750 ms, 5 s, 30 s, 200 s, and 1500 s. These sweep
times check all of the sweep time multipliers in the Sweep Generator
Current Source circuit (A22 Frequency Control). Sweep start-up time
(offset) is measured and subtracted from the measured sweep time.
The measured sweep time is compared to the selected sweep time and
the percent error calculated.
Operation Verification 5-21