Specifications
8. Line Related
Sidebands
Specification
85
dB
below the peak of a CW signal (Option 400:
~75
dB)
Related Performance
Line Related Sidebands Test
Test
Description
A signal source is connected to the spectrum analyzer input and
the necessary front-panel control settings made for the test. The
harmonics of the line frequency are calculated, the front-panel
controls to view the frequencies, and measure the amplitude of the
signal at each of the frequencies.
5-18 Operation Verification