Specifications
7. Impulse and Resolution Bandwidths (Opt. 462 Impulse Bandwidth)
7. Impulse and
Resolution
Bandwidths (Opt.
462 Impulse
Bandwidth)
Specification Bandwidth
Amplitude
3 MHz;
&20%
1 MHz to 1
kHz;
&20%
300 Hz to 10 Hz; +50%, -0%
3 MHz; fl.O
dB
1 MHz to 100 Hz; f0.5
dB
30 Hz;
f0.8
dB
10 Hz; f2.0
dB
Selectivity (60 dB/6
dB
Ratio)
3 MHz to 100
kHz;
<ll:l
30 kHz to
30 Hz;
<8:
1
10 Hz; ~100 Hz separation of 60
dB
points
Related Performance
Tests
Impulse
&
Resolution Bandwidth Accuracy Test
Impulse
&
Resolution Bandwidth Selectivity Test
Impulse
&
Resolution Bandwidth Switching Uncertainty Test
Description
A source outputting pulsed RF signals is connected to the spectrum
analyzer input. The analyzer steps through the bandwidths from
3 MHz to 1 kHz, and the controller measures the frequency of the
impulse response for each bandwidth.
Next, a CW signal is connected to the spectrum analyzer input.
The analyzer steps through all bandwidths for 3 MHz to 10 Hz,
centers the signal, sets the signal peak near the reference level, and
measures the frequency of the 6
dB
points for each bandwidth. The
6
dB
bandwidth is then calculated by determining the difference in
frequency between the 6
dB
points.
Last, the analyzer steps through the bandwidths and measures the
60
dB
points for each bandwidth. The 60 dB bandwidth is calculated
for each bandwidth. The selectivity of each bandwidth is then
calculated by dividing the 60
dB
bandwidth by the 6
dB
bandwidth
for each filter.
Note
This test is for Option 462 instruments only. For standard instruments,
perform the test on the preceding page.
Operation Verification 5-17