Specifications

6. Resolution
Bandwidths
Specification
Related Performance
Tests
Description
Note
Bandwidth
Amplitude
10 Hz to 3 MHz; *20%
3 kHz to 1 MHz;
flO%
(Referenced to 1 MHz amplitude)
3 MHz; fl.O
dB
1 MHz to 100 Hz;
f0.5
dB
30 Hz;
AO.8
dB
10 Hz; f2.0
dB
Selectivity (60
dB/3
dB
Ratio)
3 MHz to 100
kHz;
<15: 1
30 kHz to 10
kHz;
<13: 1
3 kHz to 30 Hz;
<ll:l
10 Hz; ~100 Hz separation of 60
dB
points
Resolution Bandwidth Accuracy Test
Resolution Bandwidth Selectivity Test
Resolution Bandwidth Switching Uncertainty Test
A signal source is connected to the spectrum analyzer input. The
analyzer steps through the bandwidths from 3 MHz to 10 Hz, centers
the signal, sets signal peak near the reference level, and measures
the frequency of the 3
dB
points for each bandwidth. The 3
dB
bandwidth is then calculated by determining the difference in
frequency between the 3
dB
points.
Next, the spectrum analyzer steps through the bandwidths and
measures the frequency of the 60
dB
points of each bandwidth. The
60
dB
bandwidth is then calculated by determining the frequency
difference between the 60
dB
points.
The shape factors are calculated by dividing the 60
dB
bandwidths by
the 3
dB
bandwidths.
This procedure is for instruments other than those having Option 462.
For Option 462 instruments perform the test on the following page.
5-16 Operation Verification