Service manual

Agilent 85309A H40, H41, H42 Operating and Service Manual Modification
7
Revised General Information
Absolute Maximum
Rating
Agilent Technologies
85309A H40, H41, and
H42 Operating
Characteristics
The following parameters are unique to Agilent Technologies 85309A
options H40, H41 and H42. These parameters supersede Tables 5-5, 5-6, and
5-7 in the Agilent Technologies 85310A Operating and Service Manual.
Nominal Channel
Performance
Parameter Values
LO input port (CW) + 23 dBm
Reference channel IF input port (CW) + 13 dBm
Reference channel detector input ± 20 vdc
Pos. Z blanking input ± 10 vdc
Select HI/LOW input ± 5.5 vdc
Parameter Values
Frequency range:
Low band
High band
0.1 to 1.0 GHz
0.3 to 18 GHz
Input power range (LO input recommended):
0.1 to 1.0 GHz
0.3 to 3.0 GHz
2.0 to 18 GHz
6 to 10 dBm
6 to 10 dBm
0 to 6 dBm
Power output (LO ports):
0.1 to 1.0 GHz
0.3 to 0.5 GHz
0.5 to 3.0 GHz
3.0 to 6.2 GHz
6.2 to 18 GHz
Minimum
14 dBm
21.2 dm
22.1 dBm
24.1 dBm
21.5 dBm
Typical minimum
15 dBm
24.2 dBm
24.5 dBm
26.5 dBm
24 dBm
IF channel small signal gain:
20 MHz
Minimum
21 dB
Maximum
25 dB
Output power channel tracking (typical):
0.1 to 1.0 GHz
0.3 to 3.0 GHz
2.0 to 18 GHz
± 1.3 dB
± 2 dB
± 2 dB
Port return loss, 0.1 to 18 GHz typical:
LO input
LO output
9 dB
7 dB