Agilent 85309A H40, H41, H42 Operating and Service Manual Modification
Introduction
The spurious signal levels seen on the Agilent Technologies 8510/30 depend
on the LO power level setting.
At the recommended +10 dBm setting, spur levels should be no greater
than –100 dB.
With a power level of as much as +23 dBm, spurs may be detected as
high as –60 dB.