18 Agilent 85309A H20, H21 Operating and Service Manual Modification
Revised General Information
Revised General Information
The following general information supercedes “General Information, Table
5-3”, in the Agilent 85310A Operating and Service Manual.
Typical
Agilent 85301B System
Performance Data
Other
Environmental Characteristics
Operating conditions: 0 °C to + 50 °C
Power consumption: 300 Vac (maximum)
Table 1 Typical Agilent 85301B System Performance Data with Agilent 85309A
H20 and H21
Table 5-3a
Specification (Typical) GHz *0.1 - 0.3 *0.3 - 0.8 *0.8 - 1 **0.3 - 3 2 - 3 3 - 18 ***6 - 20 ***20 - 26.5
a Sensitivity (S/N=1, 0 average) -dBm 110 110 110 110 115 115 105 100
b Compression Level (at 0.1 dB) -dBm 20 20 20 20 20 20 15 15
c Dynamic Range dB 90 90 90 90 95 95 90 85
d Channel Isolation dB 100 95 90 105 110 105 110 105
e Minimum phase lock power -dB 55 55 55 55 55 55 55 55
RF Port match (2.0:1 max) dB8888888 8
* w/85320A/B H20 mixer, Low Band; ** w/85320A/B H20, High Band; *** 3rd Harm. Mode
a Sensitivity is the calculated difference between IF noise and RF/IF conversion gain/loss. Averaging will improve sensitivity by 10 log (avg. factor).
b RF level for 0.1 dB compression: the RF input level where the RF and the IF levels are no longer tracking each other linearly within 0.1dB.
c Dynamic range is the calculated difference between 0.1 dB compression and sensitivity.
d Crosstalk is the coherent RF leakage from the reference ch3annel to the test channel with 1024 averages.
e Refers to systems that use an HP 8350 LO source. Minimum phase lock power is the minimum RF power into the reference mixer to achieve
phase lock. This does not apply to systems with a synthesized LO.