Specifications

36
PNA-X Network Analyzers
PNA-X Configuration Information
Description
Additional information
Test set
Option 200 2-ports, single source
Option 224 2-ports, add internal 2nd source,
combiner and mechanical switches
Requires Options 200, one of 219
or H85, and 080
Option 400 4-ports, dual source Option 080 recommended
Option 423 4-ports, add internal combiner and
mechanical switches
Requires Options 400, one of 419
or H85, and 080
Power configuration
Option 219 2-ports, extended powerrange and bias-tees
Option 419 4-ports, extended power range and bias-tees
Option H85
1
High power configurable (for 2- or 4-port)
Measurement applications
Option 007 Automatic fixture removal Windows 7 OS required (upgrade kit N8983A) and N52xxAU-007
Option 010 Time-domain measurements
Option 028
2
Noise figure measurements using standard receivers Requires Option 082 or 083 for
measuring frequency converters
Option 029
2
Fully-corrected noise figure measurements Requires Option 080 and for N5241/42A, one of Options 219, 224, 419, 423 or H85. For N5244/45/47A,
requires Option 224 or 423. On N5247A, noise receivers work up to 50 GHz only. For measuring frequency
converters, requires Option 082 or 083.
Option 080 Frequency offset
Option 082
3
Scalar-calibrated converter measurements Requires Option 080
Option 083
3
Vector- and scalar-calibrated converter
measurements
Requires Option 080
Option 084 Embedded LO measurements Requires at least one of Options
028, 029, 082, 083, 086, or 087
Option 086 Gain compression application
Recommend Options 219, 419 or H85 and
for measuring frequency converters, requires Option 082 or 083
Option 087 Intermodulation distortion application Requires Options 224 or 423 and for measuring frequency converters, requires Option 082 or 083
Option 088 Source phase control
Option 460 Integrated true-mode stimulus application Requires Option 400
Option 551
4
N-port capabilities
Nonlinear vector network analysis
Option 510 Nonlinear component characterization Requires Options 419 and 080,
or 400, H85 and 080
Option 514 Nonlinear X-parameters
5
Requires Options 423 and 510
Option 518 Nonlinear pulse envelope domain Requires Options 021 and 025
and either one of 510 or 514
Option 520 Arbitrary load-impedance X-parameters Requires Option 514
Required NVNA accessories
•U9391C10MHzto26.5GHzorU9391F10MHzto50GHzcombgenerator(tworequiredfornonlinearmeasurements)
•AgilentpowermeterandsensororUSBpowersensor
•Agilentcalibrationkit,mechanicalorECal
•Agilentsignalgenerator,MXGorPSGusedforX-parameterextraction(internal10MHzreferenceoutputcanbeusedfor10MHztonespacingapplications)
1. Order special model N524xAS instead of N524xA and add items N524xA-200 and
N524xAS-H85 for 2-port, extended power range, high power configuration, or items
N524xA-400 and N524xAS-H85 for 4-port, extended power range, high power configura-
tion. Order N524xA-xxx items for other standard options. Option H85 includes the extended
power range of Options 219 and 419, and therefore, they cannot be ordered together.
2. For source-corrected measurements, Options 028 and 029 on N5241/42/49A units require
an ECal module for use as an impedance tuner. N5244/45/47A units include a built-in
tuner. For calibration, Options 029 requires either a 346-series noise source (Agilent 346C
recommended) or a power meter, while Option 028 requires a power meter. All options
require a power meter for measuring mixers and converters.
3. Option 082 is a subset of Option 083; therefore, they cannot be ordered together.
4. When configured as a multiport analyzer using Option 551 and a multiport test set, the
combiner feature of Option 224 or 423 is temporarily disabled. When configured as a
standalone analyzer, the combiner feature is enabled. When ordering a test set, select an
option to specify the appropriate interconnect jumper cable set between the analyzer and
the test set.
5. X-parameters is a trademark of Agilent Technologies
Available options