G Network Analyzer Performance Specification Summary
Network Analyzer System Performance
The following specifications describe the system performance of the
Agilent E8364B PNA Series vector network analyzer Option 014 and UNL
configuration for measurements between 45 MHz and 50 GHz.
Calibration Kit Agilent 85056A, 2.4 mm precision, with sliding loads
Cables Agilent 85133F, 2.4 mm flexible test port cable set
Calibration Type Full 2- port with sliding loads
Temperatu re
Range
25°C ±5°C
Warm-up Time ≥ 0.5 hour
Maximum Output Power
Dynamic Range
System dynamic range is calculated as the difference between the noise
floor and the source maximum output power. Reflection measurements are
limited by directivity. Therefore, system dynamic range only applies to
transmission measurements.
Frequency Range (GHz)
0.045 to 10.0 10.0 to 20.0 20.0 to 30.0 30.0 to 40.0 40.0 to 45.0 45.0 to 50.0
Maximum output power +3 dBm 0 dBm −4 dBm −8 dBm −11 dBm −17 dBm
Frequency Range (GHz)
0.045 to 0.5 0.5 to 2.0 2.0 to 10.0 10.0 to 20.0 20.0 to 30.0 30.0 to 40.0 40.0 to 45.0 45.0 to 50.0
System
dynamic
range
92 dB 117 dB 120 dB 119 dB 109 dB 105 dB 102 dB 95 dB