Service manual

8510XF Network Analyzer Systems 3-23
Operation
Standard Types
Load/Short Standards These standards are used to measure reflection for:
S11 1-Port calibrations
S22 1-Port calibrations
One-Path 2-Port calibrations
Full 2-Port calibrations
When you perform any of these calibrations using a 1.0 mm calibration kit,
you will see the following menu:
Figure 3-11 Load/Short Standard Menu & Equivalent GPIB Commands
Notes on the Standards {50GHz LOAD} Refers to either of two 50 GHz loads (with male and
female connectors, as appropriate). This standard is used
for frequencies up to 50 GHz.
{(F) SHORT4} Refers to Short No. 4 with a male connector (that is, the
short is to be connected to a female port). This standard is
used for frequencies from 50 to 75 GHz.
{(F) SHORT2} Refers to Short No. 2 with a male connector (that is, the
short is to be connected to a female port). This standard is
used for frequencies above 75 GHz.
{(M) SHORT4} Refers to Short No. 4 with a female connector (that is, the
open is to be connected to a male port). This standard is
used for frequencies from 50 to 75 GHz.
{(M) SHORT2} Refers to Short No. 2 with a female connector (that is, the
short is to be connected to a male port). This standard is
used for frequencies above 75 GHz.