Service manual
8510XF Network Analyzer Systems 3-21
Operation
Standard Types
Open/Short Standards These standards are used to measure reflection for:
• S11 1-Port calibrations
• S22 1-Port calibrations
• One-Path 2-Port calibrations
• Full 2-Port calibrations
When you perform any of these calibrations using a 1.0 mm calibration kit,
you will see the following menu:
Figure 3-9 Open/Short Standard Menu & Equivalent GPIB Commands
Notes on the Standards {(F) OPEN} Refers to an open which has a male connector (that is, the
open is to be connected to a female port). This standard is
used for frequencies up to 50 GHz.
{(F) SHORT3} Refers to Short No. 3 with a male connector (that is, the
short is to be connected to a female port). This standard is
used for frequencies above 50 GHz.
{(M) OPEN} Refers to an open which has a female connector (that is,
the open is to be connected to a male port). This standard
is used for frequencies up to 50 GHz.
{(M) SHORT3} Refers to Short No. 3 with a female connector (that is, the
short is to be connected to a male port). This standard is
used for frequencies above 50 GHz.