Service manual
8510XF Network Analyzer Systems 3-15
Operation
Calibration Types
Response & Isolation This is a very limited calibration, which corrects only for frequency response
and isolation, not for directivity or source match errors. The performance of
the 8510XF is not specified, if this calibration is used. To perform this
calibration using the 1.0 mm calibration kit, select:
[CAL] {CAL1 or CAL2} {RESPONSE & ISOL’N}
This will call up the menu and submenu shown below:
Figure 3-4 Response & Isolation Menu, Response Menu, and Equivalent GPIB
Commands
Select ONE of the
Response Standards
For Reflection Measurements (S11, S22), select one of the following:
{(F) SHORT3} Refers to offset short No. 3 with a male connector (that
is, the short is to be connected to a female port).
{(F) OPEN} Refers to an open with a male connector (that is, the open
is to be connected to a female port).
{(M) SHORT3} Refers to offset short No. 3 with a female connector (that
is, the short is to be connected to a male port).
{(M) OPEN} Refers to an open with a female connector (that is, the
open is to be connected to a male port).
For Transmission Measurements (S12, S21), select:
{THRU} Connect Port 1 to Port 2 by way of a cable.
Isolation Standard {ISOL’N STD
LOAD}
Refer to “Isolation” under “Banded Standards” in
Table 3-1 on page 3-12.