3-14 8510XF Network Analyzer Systems
Operation
Calibration Types
Response This is a very limited calibration, which corrects only for frequency
response, not for directivity or source match errors. The performance of the
8510XF is not specified, if this calibration is used. To perform this
calibration using the 1.0 mm calibration kit, select:
[CAL] {CAL1 or CAL2} {CALIBRATE: RESPONSE}
This calls up the response standard select menu, as shown on the right of the
illustration below:
Figure 3-3 Response Menu (Right) & Equivalent GPIB Commands
Select ONE of the
Response Standards
For Reflection Measurements (S11, S22), select one of the following:
{(F) SHORT3} Refers to offset short No. 3 with a male connector (that
is, the short is to be connected to a female port).
{(F) OPEN} Refers to an open with a male connector (that is, the open
is to be connected to a female port).
{(M) SHORT3} Refers to offset short No. 3 with a female connector (that
is, the short is to be connected to a male port).
{(M) OPEN} Refers to an open with a female connector (that is, the
open is to be connected to a male port).
For Transmission Measurements (S12, S21), select:
{THRU} Connect Port 1 to Port 2 by way of a cable.