Service manual
3-4 8510XF Network Analyzer Systems
Operation
8510XF Operating System
Menu Changes
(Overview)
The [CAL], [DOMAIN], [PARAMETER], STIMULUS[MENU], and [SYSTEM] menus have all
been modified for the 8510XF, as noted briefly below.
CAL Menu The [CAL] menu is changed in the following areas:
• {TRIM SWEEP} is deleted.
• The contents of the [CAL] menu (and most of its submenus) reflect the
name of the calibration kit that is in use, as well as the names of the
calibration standards in that kit. A calibration kit (85059A) has been
developed specifically for the 8510XF; the menu maps in Chapter 7
reflect the contents of that kit.
The
[CAL] menu and its submenus are discussed later in this chapter (see
“Calibration Types” on page 3-13).
DOMAIN Menu The [DOMAIN] menu is changed as follows:
{PULSE PROFILE} is deleted.
PARAMETER Menu The [PARAMETER] menu is changed as follows:
{PHASE LOCK} is deleted.
STIMULUS Menu STIMULUS [MENU] is changed in the following areas:
• The submenu called up by the {POWER} softkey is substantially revised.
{ATTENUATOR PORT 1}, {ATTENUATOR PORT 2}, and {POWER FLATNESS} are
deleted. Also, new port power functions are added (see “Port Power” on
page 3-28).
• {RAMP} is deleted.
• {COUPLED CHANNELS} and {UNCOUPLED CHANNELS} are deleted.