8510XF Network Analyzer Systems 2-37
Installation
Cable List
Other Connections
and Settings
GPIB Addresses When the 8510C power is turned on, all previously assigned GPIB addresses
are automatically recalled from memory and assigned to the
various system instruments, including the address of the 8510C itself.
The GPIB address switch settings for all instruments must match the
addresses assigned by the system. In 8510XF systems, these addresses are
set at the factory prior to shipment. For reference, the instruments and the
applicable default addresses are listed in the table below.
Language Switches The LANG switches of the 83621B and 83651B sources must be set to 111.
Table 2-13 8510XF System GPIB Addresses
Instrument GPIB Address
8510C 16
System Bus 17
83621B LO Source 18
83651B RF Source 19
Millimeter-wave controller 23
Plotter 5
Printer 1
Disk Drive 0