Service manual

2-10 8510XF Network Analyzer Systems
Installation
Precautions
WARNING If this product is not used as specified, the protection provided by the
equipmentcouldbeimpaired. Thisproductmustbeusedinanormalcondition
(in which all means for protection are intact) only.
Electrostatic
Discharge
CAUTION The millimeter-wave controller and the test heads are sensitive to
electrostatic discharge (ESD). Ground your work station before unpacking
and installing the test heads. See “Electrostatic Discharge” on page 5-2 for
more information.
System Voltages All instruments in the 8510XF system must be set to the same voltage as the
system rack (either 120 Vac or 220 Vac). 120 Vac is the factory setting.
CAUTION The cabinet fans may be permanently damaged if a 120V system is plugged
into a 220V ac power outlet. The cabinet fans can be wired for either 120V
or 220V, but not both. Therefore, a system wired for 120V operation cannot
be switched to 230V operation simply by changing the voltage selection
switches on individual instruments.
Test Port Inputs
CAUTION Input power to the test ports must not exceed +27 dBm. Input power in
excess of this level will damage expensive components. Observe proper
precautions, especially when measuring amplifiers with gains of 20 dB or
greater.
Cleaning
WARNING To prevent electrical shock, disconnect the E734X/E735X mains before
cleaning. Use a dry cloth or one slightly dampened with water to clear the
external case parts. Do not attempt to clean internally.