Service manual
8510XF Network Analyzer Systems 2-9
Installation
Precautions
Precautions
Safe Installation Install the system so that the ON/OFF switch is readily identifiable and is
easily reached by the operator. The ON/OFF switch or the detachable power
cord is the system disconnecting device. It disconnects the mains circuits
from the mains supply before other parts of the system. Alternatively, an
externally installed switch or circuit breaker (which is readily identifiable
and is easily reached by the operator) may be used as a disconnecting device.
Install the system according to the enclosure protection provided:
• The system protects against finger access to hazardous parts within the
enclosure.
• The system does not protect against the ingress of water.
CAUTION This product is designed for use in Installation Category II and Pollution
Degree 2 per IEC 1010 and 664 respectively.
CAUTION Always use the three-prong ac power cord supplied with this product.
Failure to insure adequate earth grounding by not using this cord may cause
product damage.
CAUTION Before switching on the system, make sure that the correct fuse is installed,
and that the supply voltage is in the specified range.
CAUTION Ventilation requirements:
Convection in and out of the system cabinet must not be restricted. The
ambient temperature (outside the cabinet) must be less than the maximum
operating temperature of the system by 4°C for every 100 watts dissipated in
the cabinet. If the total power dissipated in the cabinet is greater than 800
watts, then forced convection must be used.
WARNING This is a Safety Class 1 Product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall only be inserted in a
socket outlet provided with a protective earth contact. Any interruption of the
protective conductor inside or outside the product is likely to make the
product dangerous. Intentional interruption is prohibited.