Service manual
7-46 8510XF Network Analyzer Systems
Menus & Commands
New GPIB Commands
SOU2OFF [SYSTEM] {MORE} {LO POWER CONFIG}
{LO LEVEL -- LEVELING OFF}
Places the 83621B LO source in leveling-off
mode.
SOU2POWOFF [SYSTEM] {MORE} {LO POWER CONFIG}
{LO POWER OFF}
Deactivates RF power from the 83621B LO
source.
SOU2POWON [SYSTEM] {MORE} {LO POWER CONFIG}
{LO POWER ON}
Activates RF power from the 83621B LO
source.
SOU2POW? No softkey equivalent Queries the status of LO power (0 for OFF,
1 for ON).
SOU2SYS [SYSTEM] {MORE} {LO POWER CONFIG}
{LO LEVEL -- SYSTEM}
Places leveling of the 83621B LO source
under control of the 8510XF.
TSPEEK [SYSTEM] {MORE} {SERVICE}
{XF TESTSET SERVICE} {PEEK TEST
SET}
Used by Agilent service engineers to read
from a location (specified by the TSPEEL
command) in the memory of the
millimeter- wave controller.
TSPEEL [SYSTEM] {MORE} {SERVICE}
{XF TESTSET SERVICE} {PEEK/POKE
LOCATION}
Used byAgilent service engineers to specify
a location in the memory of the
millimeter-wave controller.
TSPOKE <byte> [SYSTEM] {MORE} {SERVICE}
{XF TESTSET SERVICE} {POKE TEST
SET}
Used by Agilent service engineers to write a
byte of data to a location (specified by the
TSPEEL command) in the memory of the
millimeter-wave controller.
COMMAND EQUIVALENT SOFTKEY FUNCTION