8510XF Network Analyzer Systems 5-5
System Maintenance
1.0 mm Connector Care
If a connector appears defective, clean and inspect it again. Damaged
connectors should be discarded or sent for repair. Try to determine the cause
of damage before connecting a new, undamaged connector in the same
configuration.
Cleaning connectors
1. Inspect the connectors for dirt or debris.
2. Blow off any accumulated dust, using clean (filtered) compressed air.
3. Dip the cleaning swab in isopropyl alcohol.
NOTE Do not pour alcohol directly on the parts to be cleaned, and do not allow
alcohol to wash down inside the connectors. An excessive flow of alcohol
may carry residue into the interior of the connectors.
4. Gently wipe all connecting surfaces with the end of the cleaning swab.
5. Dry with compressed air.
6. Inspect the connectors again, and repeat the cleaning procedure if
necessary.
Connector care supplies
The following supplies can be ordered from Agilent Technologies:
Item Description Part Number
Magnifying lens (10x) 1000-1114
(included in Calibration Kit)
Isopropyl Alcohol (30 ml, 99.5%) 8500-5344
Cleaning Swabs (100 each) 9301-1243
Torque Wrench (6 mm, 4 in-lb) 8710-2079 (included in Calibration Kit)
Open End Wrench (6 mm) 8710-2156 (included in Calibration Kit)