Service manual
The 8510XF requires separate verifications for frequencies below
and above 50 GHz (see "Frequency Ranges" in this same chapter).
A special two-band calibration procedure is required for the
performance verification of the 8510XF System (a single full-band
measurement calibration is equivalent).not
There are up to three different full 2-port calibrations and
corresponding frequency ranges required, which are stored
in suggested locations in the 8510 analyzer.
Two-Band Calibration Procedure for
8510XF Performance Verification
8510XF Full 2-Port Calibration Menus
Calibration Techniques and Standards
Calibration Kit Label:
Disk File Name:
1.00mmA.1A
CK_MMKA1A
1. SOLT = Short, Open, Load and Thru standards
2. The broadband load is a combination of a lossy delay line plus a 50 GHz load.
3. Also used for 45 MHz frequency point.
Frequency
Range
DC–50 GHz
3
Calibration
Technique
SOLT
1
"Open" Standard
(Reflection)
Open
"Short" Standard
(Reflection)
Short 3
"Load" Standard
(Reflection)
50 GHz Load Thru
"Isolation"
Standard
"Thru" Standard
(not used for
1-port cal)
50 GHz Load
50–110 GHz Offset Shorts Short 3 Short 1
Short 4
Short 2
Thru Load BB
2
Calibration Storage Locations
1. This calibration is for Option 005 systems only.
2. The frequency for standard systems is 2–50 GHz.
3. The frequency range shown is for the E7352A system only.
The E7342A system's frequency range will be: 50–85 GHz
File Location Name:
Cal Set 1
Frequency Range of Data Stored:
1–50 GHz (2–50 GHz)
12
Cal Set 2 45 MHz
1
Cal Set 3 50–110 GHz
3
Calibration Techniques:
A variety of calibration techniques can be used with the 8510XF, including the following:
(Short-Open-Load-Thru) Used for performance verification 45 MHz to 50 GHz.
(Shorts with different offset lengths) Used for performance verification 50 GHz
to 110 GHz.
(Thru-Reflect-Line)**
(Thru-Reflect-Match)**
** Note: Not used for performance verification
The calibration technique used depends in part on the type of calibration kit used. The 1.0 mm
calibration kit (85059A) was designed specifically for the 8510XF, and uses a combination of
two calibration techniques: SOLT calibration for frequencies up to 50 GHz, and offset shorts
calibration for frequencies above 50 GHz. This combination of techniques provides the best
possible calibration results on the 8510XF.
•
SOLT
Offset Shorts
TRL
TRM
•
•
•
rr51c
50 GHz Load
Lossy Delay Line
"FWD/REV ISOL'N LOAD"
For: > 50 GHz a lossy delay line is required in addition
to a 50 GHz load.*
For: < 50 GHz use a 50 GHz load only.
Note: If you use a 50 GHz load only, a bad calibration will
result due to reflection from the load above 50 GHz.
*
REFLECT'N
TRANS-
MISSION
ISOLATION
SAVE
2-PORT CAL
REFL
TRAN
ISOL
SAV2
FWDT
CLASS11B
CLASS22C
CLASS22A
REFD
CLASS11A
CLASS22B
CLASS11C
FWDM
OMII
REVT
REVM
FWDI
REVI
TRAD
ISOD
Full 2-Port
Cal Menu
Cal
Cal Set
Select Menu
(S11):
OPEN/SHORT
SHORTS
LOAD/SHORT
(S22):
OPEN/SHORT
SHORTS
LOAD/SHORT
REFLECT'N
DONE
FWD. TRANS
THRU
FWD. MATCH
THRU
REV. TRANS.
THRU
REV. MATCH
THRU
TRANS.
DONE
OMIT
ISOLATION
FWD ISOL'N
LOAD
REV ISOL'N
LOAD
ISOLATION
DONE
Full 2-Port
Reflection
Cal Menu
Full 2-Port
Transmission
Cal Menu
Full 2-Port
Isolation
Cal Menu
(F) OPEN
(F) SHORT3
(M) OPEN
(M) SHORT3
DONE
OPEN/SHORT
(F) SHORT3
(F) SHORT1
(M) SHORT3
(M) SHORT1
DONE
SHORTS
50 GHz LOAD
(F) SHORT4
(F) SHORT2
(M) SHORT4
(M) SHORT2
DONE
LOAD/SHORT
Open/Short
Shorts Menu
Load/Short
Menu
(F)
(M)
(M)
(M)
(F)
(F)
(F) (M)
(M)
(F)
(M) (F)
(M)
(M)
(M)
(F)
(F)
(M)
(F)
(F)
DONE
DONE
DONE
Right
Test
Head
(E7352R or
E7342R)
Left
Test
Head
(E7352L or
E7342L)
E7341A
(Std. or
Opt. 005)
Cable
F-F (8.8 cm)
Full 2-port calibration system configuration
for performance verification in 1.0 mm coax.
Cable F-F (8.8 cm) 11500-60001
From 85059A calibration verification kit
83651B
83621B
MM Controller
8510C
s1
F
or more information refer to the section on "Frequency Ranges"
in this same chapter. The parameters for these calibrations are set up
automatically on the analyzer by the performance verification software.
Clarifying Connector Sex:
The calibration standard labels that appear in the 8510XF softkey
calibration menus specify connector sex as "(M)" or "(F)". The sex designator
refers to the sex of the test port connector to which the calibration standard
is connected ( the sex of the calibration standard connector).not
Offset Shorts:
Offset shorts are used in the place of opens and loads
The "OPEN" category is therefore renamed "OPEN/SHORT," and the "LOAD"
category is renamed "LOAD/SHORT" on the calibration menu softkey labels.
at frequencies above 50 GHz.
Male Test Port