Service manual
4-16
8510XF Network Analyzer Systems
Performance Verification
CW Frequency Accuracy Test
Performance Test
Record
Record the measured frequencies for RF source readings on Table 4-7, and
for LO source on Table 4-8.
Table 4-7 Performance Test Record for CW Frequency Accuracy Test (RF Source)
Instrument Model: _________________ Report Number: __________________________ Date: _______
Frequency Minimum Specification Recorded Results Maximum Specification Uncertainty
1
45 MHz 44.999955 MHz 45.000045 MHz 10 Hz
2 MHz 1.999998 GHz 2.000002 GHz 10 Hz
20 GHz
2
19.99998 GHz 20.00002 GHz 4 kHz
26 GHz 26.4999735 GHz 26.5000256 GHz 5 kHz
1. The measurement uncertainty is quoted for these performance tests using only the recommended models specified in Table 8-1 of the “Equipment
Required” chapter of the On-Site Service Manual. The quoted uncertainty represents limits of 3 times the equivalent standard deviation (3s) and is
intended to represent 90% confidence level.
2. For 83620A/B or 83621A/B only.
Table 4-8 Performance Test Record for CW Frequency Accuracy Test (LO Source)
Instrument Model: _________________ Report Number: __________________________ Date: _______
Frequency Minimum Specification Recorded Results Maximum Specification Uncertainty
1
45 MHz 44.999955 MHz 45.000045 MHz 10 Hz
2 MHz 1.999998 GHz 2.000002 GHz 10 Hz
20 GHz
2
19.99998 GHz 20.00002 GHz 4 kHz
1. The measurement uncertainty is quoted for these performance tests using only the recommended models specified in Table 8-1 of the “Equipment
Required” chapter of the On-Site Service Manual. The quoted uncertainty represents limits of 3 times the equivalent standard deviation (3s) and is
intended to represent 90% confidence level.
2. For 83620A/B or 83621A/B only.