Service manual
4-10
8510XF Network Analyzer Systems
Performance Verification
Verification Procedures
c. To make the thru connection, gently slide the right test (port 2)
towards the left test head (port 1) a little bit at a time, while turning
the threaded ring on the left head test port connector by hand onto
the cable attached to the right test head. Do not use the threaded ring
on the test port connector to pull the cable into the connector. Repeat
this process until the cable is firmly seated into port 1, then make the
threaded ring barely finger tight. Finally use the torque wrench on
the port 1 threaded ring and a backup wrench on the cable to tighten
the connection.
4. Select
Select Standard to go to the Standard Selection Menu.
5. To verify the system use the Verification Standards provided with the
85059A 1.0 mm Precision Calibration and Verification Kit (refer to
Table 4-6). Using the menu select the verification standard and enter the
standard’s serial number. The verification standards may be measured in
any order.
NOTE It is recommended to connect the Verification Standards to the cable on
port 2 prior to attempting to connect to port 1. Then follow the
recommended procedure for making a thru connection outlined under step 3
above to avoid damaging the cable.
6. Use defaults for
Cal Set for Broadband Measurement: 1 and Cal Set for 45 MHz
Measurement (If Required): 2
in the Standard Menu.
7. If your system has Option 005, verify that
Cal Set for 45 MHz Measurement (If
Required)
is set to Yes. Otherwise, set to No.
8. Select
Done and follow the program prompts.
9. Select
Measure Data to measure the verification standards. Follow the
program prompts.
10. When verification has been completed view the data, then print or save
(if required). If needed the verification standard can be remeasured by
selecting
Repeat Measure.
Table 4-6 Verification Standards from 1.0 mm Calibration and Verification Kit
Type of Standard Part Number
Mismatched Thru Adapter 85059-60016
Match Thru Adapter 85059-60017