Service manual

8510XF Network Analyzer Systems
4-3
Performance Verification
Materials Required The following materials are required to run the tests:
8510XF system with accessories
85059A 1.0 mm Precision Calibration and Verification Kit
PC-based, 8510 Specification and Performance Verification Software
(Rev. A.05.00 or greater; p/n 08510-10033)
Laptop or PC running BASIC for Windows (3.1/95/NT)
(Rev. 6.3 or greater)
GPIB Card for PCs (National Instruments or HP)
PCMCIA Card for Omnibook or laptop
GPIB cable from PC to 8510XF
Calibration and
Frequency Ranges
The 8510XF requires separate verifications for frequencies below and above
50 GHz (see on “Frequency ranges” on page 4-4).
A special two-band calibration procedure is required for the performance
verification of the 8510XF system (a measurement calibration is not
sufficient). There are up to three different calibrations and corresponding
frequency ranges required, which are stored in suggested locations in the
8510 analyzer (Refer to Table 4-1 on page 4-3). For more information refer
to the section on “Frequency ranges” on page 4-4. The parameters for these
calibrations are set up automatically on the analyzer by the performance
verification software.
Table 4-1 Calibration Storage Locations
File Location Name:
Frequency Range of Data
Stored:
Cal Set 1 1-50 GHz
1
(2-50 GHz)
2
1. This calibration is for Option 005 systems only.
2. The frequency for standard systems is 2-50 GHz.
Cal Set 2 45 MHz
1
Cal Set 3 50-110 GHz
3
3. The frequency range shown is for the E7352A system only. The E7342A
system’s frequency range will be: 50-85 GHz.