Service manual

8510XF Network Analyzer Systems 3-37
Operation
LO Power
If {LO LEVEL / EXTERNAL} is selected, the LO source performs its own leveling,
using an external detector.
If
{LO LEVEL / LEVELING OFF} is selected, the LO source is set to the unleveled
mode.
LO Power Settings The {LO POWER SETTINGS} softkey calls up a submenu (see “LO Power
Settings” below).
LO Power On/Off The softkeys {LO POWER /ON} and {LO POWER /OFF} activate and deactivate LO
source power.
LO Power Settings [SYSTEM] {MORE} {LO POWER CONFIG} {LO POWER SETTINGS} calls up the submenu
which is shown on the right of Figure 3-20 on the previous page.
NOTE The functions represented on this submenu will not work if {LO LEVEL /
SYSTEM}
is selected on the [SYSTEM] {MORE} {LO POWER CONFIG} menu (see “LO
Power Configuration” on page 3-36). These functions are not used during
normal operation of the system; they are made available for service use, or
for unusual measurement situations.
LO Source Power {LO SOURCE POWER}, followed by a number, specifies a power level setting for
the 83621B LO source, in dBm.
LO Source Slope {SLOPEON} and {SLOPE OFF} are used to enable and disable power slope for the
LO source.
{SLOPE ON}, followed by a number, specifies the power slope
setting for the LO source, in dB/GHz.