Service manual
2 Agilent 8494A/B, 8495A/B, 8496A/B Operating And Service Manual
Description
Description
Agilent 8494A/B, 8495A/B, and 8496A/B are 50-ohm coaxial step
attenuators.
For the 8494A/B, the attenuation can be varied in 1 dB steps, or 10 dB steps
for the 8495A/B and 8496A/B. The attenuation shown on the control knob is
the additional attenuation added in the signal path over the insertion loss of
the attenuator in the 0 dB position.
• Agilent 8494A/B instruments are four-section attenuators with a range
of 0 dB to 11 dB in 1 dB steps.
• The 8495A/B is a three-section attenuator with a range of 0 dB to 70 dB
in 10 dB steps.
• The 8496A/B is a four-section attenuator with a range of 0 dB to 110 dB
in 10 dB steps.
The attenuator sections are connected in cascade. Each section consists of a
precision, thin-film attenuator card, a lossless thru-line and a ganged pair of
edge line transmission lines. The edge lines are flexed to make contact with
either the attenuator card or the thru-line. The edge line contacts are
gold-plated leaf springs which ensure long life and high repeatability.
Low-torque cams flex the edge lines. Table 1, “Attenuator Switching Order”
on page 7 shows the switching arrangements.
CAUTION Do not exceed the RF power rating of 1 W average or 100 W peak with a
maximum pulse width of 10
µs. Do not connect an attenuator RF input or
output connector to greater the 7Vdc. If the attenuator must be connected to
a device with a potential greater than 7 Vdc, use a blocking capacitor.
Instrument Options Each instrument is specified with an option number which denotes the
configuration of the input and output connectors.
Option Connector Description
001 Both connectors Type N female
002 Both connectors SMA female
003
1
1. Option 003 is not available with the 8494A, 8495A, and 8496A.
Both connectors APC-7