User`s guide
Table Of Contents
- Legal Information
- Equipment Operation
- General Safety Considerations
- Regulatory Information
- List of Related Publications
- Agilent TechnologiesE4418B Options
- About this Guide
- Getting Started
- Power Meter Operation
- Introduction
- Battery Operation (Option 001)
- Zeroing and Calibrating the Power Meter
- Making Measurements with the AgilentE-Series Power Sensors
- Making Measurements with the Agilent8480 Series Power Sensors
- Making Measurements using Sensor Calibration Tables
- Making Measurements using Frequency Dependent Offset Tables
- Setting the Units of Measurement
- Selecting Units of Measurement from the Softkeys
- Making Relative Measurements
- Setting the Resolution
- Setting Offsets
- Setting Averaging
- Measuring Pulsed Signals
- Setting Measurement Limits
- Selecting a Digital or Analog Display
- Setting the Range
- Configuring the Remote Interface
- Recorder Output
- Saving and Recalling Power Meter Configurations
- How Measurements are Calculated
- Presetting the Power Meter
- Self Test
- Operator Maintenance
- Contacting Agilent Technologies
- Menu Reference
- Error Messages
- Specifications

Power Meter Operation
Setting Measurement Limits
2-44 Agilent E4418B User’s Guide
Figure 2-11: Limits Checking Application
Figure 2-12: Limits Checking Results
In this application a swept frequency signal is applied to the input of the
Device Under Test. The power meter measures the output power. The
limits have been set at +4 dBm and +10 dBm. A fail occurs each time the
output power is outside these limits as shown in Figure 2-12.
Setting Window Limits
The power meter can be configured to verify the current measurement in
either window against predefined upper and/or lower limit values. The
range of values that can be set for the upper and lower limits and the
Power Meter
Swept Source
CHANNEL A
INPUT
OUT
IN
OUT
Device
Under Test
4 dBm
+
10 dBm
Amplitude
Frequency
o
o
o
o
o
o
o
Fail
Fail