User`s guide
Table Of Contents
- Legal Information
- Equipment Operation
- General Safety Considerations
- Regulatory Information
- List of Related Publications
- Agilent TechnologiesE4418B Options
- About this Guide
- Getting Started
- Power Meter Operation
- Introduction
- Battery Operation (Option 001)
- Zeroing and Calibrating the Power Meter
- Making Measurements with the AgilentE-Series Power Sensors
- Making Measurements with the Agilent8480 Series Power Sensors
- Making Measurements using Sensor Calibration Tables
- Making Measurements using Frequency Dependent Offset Tables
- Setting the Units of Measurement
- Selecting Units of Measurement from the Softkeys
- Making Relative Measurements
- Setting the Resolution
- Setting Offsets
- Setting Averaging
- Measuring Pulsed Signals
- Setting Measurement Limits
- Selecting a Digital or Analog Display
- Setting the Range
- Configuring the Remote Interface
- Recorder Output
- Saving and Recalling Power Meter Configurations
- How Measurements are Calculated
- Presetting the Power Meter
- Self Test
- Operator Maintenance
- Contacting Agilent Technologies
- Menu Reference
- Error Messages
- Specifications

Power Meter Operation
Setting Averaging
2-38 Agilent E4418B User’s Guide
Setting Averaging
The power meter uses a digital filter to average power readings. The
number of readings averaged can range from 1 to 1024. This filter is used
to reduce noise, obtain the desired resolution and to reduce the jitter in
the measurement results. Increasing the value of the filter length reduces
measurement noise. However, the time to take the measurement is
increased. You can select the filter length or you can set the power meter
to auto filter mode. The default is “AUTO”.
When the auto filter mode is enabled, the power meter automatically sets
the number of readings averaged together to satisfy the filtering
requirements for most power measurements. The number of readings
averaged together depends on the resolution and the power level currently
being measured. Figure 2-9 lists the number of readings averaged for each
range and resolution when the power meter is in auto filter mode and is
set to normal speed mode (refer to the Agilent E4418B/E4419B
Programming Guide for details of the readings averaged in the other
speed modes).
Resolution is a window function and not a channel function. In the case
where a channel is set up in both the upper and lower window and the
resolution settings are different, the highest resolution setting is taken to
calculate the averaging number. For example, if the upper window has a
resolution setting of 2 and is measuring channel A and the lower window
has a resolution of 4 and is also measuring channel A. In this instance
channel A averaging is calculated with a resolution of 4.