User`s guide

Table Of Contents
Power Meter Operation
Making Measurements using Sensor Calibration Tables
2-22 Agilent E4418B Users Guide
The power meter is shipped with a set of predefined sensor calibration
tables. The data in these tables is based on statistical averages for a range
of Agilent Technologies power sensors.
These power sensors are:
DEFAULT
1
Agilent 8481A
Agilent 8482A
2
Agilent 8483A
Agilent 8481D
Agilent 8485A
Agilent R8486A
Agilent Q8486A
Agilent R8486D
Agilent 8487A
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9 which do not contain any data when the power meter is
shipped from the factory.
You cannot delete any of the 20 sensor calibration tables. Each sensor
calibration table can contain a maximum of 80 frequency points each. If
you want a new sensor calibration table you must edit an existing one.
To edit a table:
1. Select one using the and hardkeys to scroll through the
displayed tables. When the table you want to edit is highlighted
press . The Edit Cal screen is displayed as shown
in Figure 2-5 on page 2-21. In this screen you can modify, add and
remove frequency and calibration factors and also edit the table
name. Use the , , and hardkeys to move
between the table name and its frequency and calibration factors.
A frequency in the range of 0.001 MHz to 999.999 GHz can be
entered.
1. DEFAULT is a sensor calibration table in which the reference calibration
factor and calibration factors are 100%. This sensor calibration table can be
used during the performance testing of the power meter.
2. The Agilent 8482B and Agilent 8482H power sensors use the same data as
the Agilent 8482A.
Edit Table
HP4402.book Page 22 Wednesday, November 29, 2000 4:46 PM