User`s guide
Table Of Contents
- Legal Information
- Equipment Operation
- General Safety Considerations
- Regulatory Information
- List of Related Publications
- Agilent TechnologiesE4418B Options
- About this Guide
- Getting Started
- Power Meter Operation
- Introduction
- Battery Operation (Option 001)
- Zeroing and Calibrating the Power Meter
- Making Measurements with the AgilentE-Series Power Sensors
- Making Measurements with the Agilent8480 Series Power Sensors
- Making Measurements using Sensor Calibration Tables
- Making Measurements using Frequency Dependent Offset Tables
- Setting the Units of Measurement
- Selecting Units of Measurement from the Softkeys
- Making Relative Measurements
- Setting the Resolution
- Setting Offsets
- Setting Averaging
- Measuring Pulsed Signals
- Setting Measurement Limits
- Selecting a Digital or Analog Display
- Setting the Range
- Configuring the Remote Interface
- Recorder Output
- Saving and Recalling Power Meter Configurations
- How Measurements are Calculated
- Presetting the Power Meter
- Self Test
- Operator Maintenance
- Contacting Agilent Technologies
- Menu Reference
- Error Messages
- Specifications

Menu Reference
Front Panel Menu Reference "Frequency/Cal Fac" Menu
3-14 Agilent E4418B User’s Guide
•
Press this softkey to enter the measurement calibration factor for
channel A in the range of 1% to 150%. The default value is 100%.
Use the , , and hardkeys to change the
calibration factor. To confirm your choice press . Entering a
calibration factor using this softkey is only possible when you are
using an Agilent 8480 series power sensor.
Cal Fac
%