Product specifications
9
E-series power sensor
specifications
The E-series of power sensors have
their calibration factors stored in
EEPROM and operate over a wide
dynamic range. They are designed
for use with the EPM series of power
meters and two classes of sensors
are available:
• CW power sensors (E4412A and
E4413A).
• Average power sensors (E9300
sensors).
Power linearity
Table 7
Power Temperature (25ºC ±5ºC) Temperature (0º to 55ºC)
100 pW to 10 mW ±3 % ±7 %
(-70 dBm to +10 dBm)
10 mW to 100 mW ±4.5% ±10%
(+10 dBm to +20 dBm)
E-series CW power sensor specifications
Wide dynamic range CW sensors: 100 pW to 100 mW (-70 dBm to +20 dBm)
Table 6
Model Frequency range Maximum SWR Maximum power Connector type
E4412A 10 MHz – 18 GHz *10 MHz to 30 MHz: 1.12 200 mW (+23 dBm) Type-N (m)
30 MHz to 2 Ghz:1.15
2 GHz to 6 Ghz:1.17
6 GHz to 11 Ghz:1.2
11 GHz to 18 Ghz:1.27
E4413A 50 MHz – 26.5 GHz 50 MHz to 100 MHz: 1.21 200 mW (+23 dBm) APC-3.5mm (m)
100 MHz to 8 Ghz:1.19
8 GHz to 18 Ghz:1.21
18 GHz to 26.5 Ghz:1.26
* Applies to sensors with serial prefix US 3848 or greater
Figure 1 : Relative mode power measurement
linearity with EPM series power meter/E-series
CW power sensor at 25°C ± 5°C (typical)