Product specifications
15
E-series E9300 average power sensors
Typical E9300B/01B power linearity at 25°C, after zero and calibration, with associated
measurement uncertainty
Power range -6 to 0 dBm 0 to +10 dBm +10 to +20 dBm +20 to +26 dBm
Measurement uncertainty ± 0.65% ± 0.55% ± 0.45% ± 0.31%
Typical E9300H/01H power linearity at 25°C, after zero and calibration, with associated
measurement uncertainty
Power range -26 to -20 dBm -20 to –10 dBm -10 to 0 dBm 0 to +10 dBm +10 to +20 dBm +20 to +26 dBm
Measurement uncertainty ± 0.9% ± 0.8% ± 0.65% ± 0.55% ± 0.45% ± 0.31%