E-series E9300 average power sensors
Table 10. Power linearity (after zero and cal at ambient environmental conditions)
Sensor Power Linearity Linearity
(25°C ± 10°C) (0 – 55°C)
E9300A, E9301A, E9304A -60 to –10 dBm ±3.0% ±3.5%
-10 to 0 dBm ±2.5% ±3.0%
0 to +20 dBm ±2.0% ±2.5%
E9300B, E9301B -30 to +20 dBm ±3.5% ±4.0%
+20 to +30 dBm ±3.0% ±3.5%
+30 to +44 dBm ±2.5% ±3.0%
E9300H, E9301H -50 to 0 dBm ±4.0% ±5.0%
0 to +10 dBm ±3.5% ±4.0%
+10 to +30 dBm ±3.0% ±3.5%
Typical E9300A/01A/04A Power Linearity at 25°C, after zero and calibration, with associated
measurement uncertainty
Power range -30 to -20 dBm -20 to -10 dBm -10 to 0 dBm 0 to +10 dBm +10 to +20 dBm
Measurement uncertainty ±0.9% ±0.8% ±0.65% ±0.55% ±0.45%