Specifications
4
Agilent 8510E Option 005
45 MHz to 20 GHz
The following specifications describe the system
performance for the 8510C network analyzer in the 8510E
Option 005 (replaces 85052D fixed loads with 85052B
sliding loads calibration kit) configuration. The system
hardware includes the following:
Test set: 8514B S-parameter test set
RF source: 83621B synthesized sweeper
Calibration kit: 85052B 3.5mm calibration kit
Calibration technique: Full two-port calibration with
sliding loads.
Dynamic range (for transmission measurements)
Frequency range (GHz)
0.045–2 2–8 8–20
Maximum power
measured at port 2 +20 dBm +11 dBm +10 dBm
Reference power
at port 1 (nominal) +2 dBm –2 dBm –6 dBm
Minimum power
measured at port 2 –66 dBm –95 dBm –95 dBm
Receiver dynamic range 86 dB 106 dB 105 dB
System dynamic range 68 dB 93 dB 89 dB
Measurement uncertainty
Frequency range (GHz)
RESIDUAL 0.045–2 2–8 8–20
Directivity 48 dB 44 dB 44 dB
Source match 40 dB 33 dB 31 dB
Load match 48 dB 44 dB 44 dB
Reflection tracking ±0.003 dB ±0.003 dB ±0.006 dB
Transmission tracking ±0.017 dB ±0.044 dB ±0.084 dB
Crosstalk 89 dB 115 dB 110 dB
Frequency range (GHz)
RAW (Typical) 0.045–2 2–8 8–20
Directivity 23 dB 23 dB 14 dB
Source match 17 dB 15 dB 11 dB
Load match 17 dB 15 dB 11 dB
Reflection measurements
Transmission measurements
System dynamic range
Magnitude Phase
Magnitude Phase
System performance