Specifications

Overload Detector DAC Adjustment
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Overload Detector DAC Adjustment
The overload detector is in the first IF in the A20 Low Band. It is used as a warning when
a high input level outside the viewing span is compressing the first mixer. It can also be
used as a diagnostic tool.
This adjustment should be performed after replacing the A20 assembly.
Required Test Equipment
Test Equipment Model Number
E4440A
E4443A, E4445A
E4446A
E4448A
Signal Generator E4433B X X
Type N Cable 11500C X X
Type N (f) to 3.5 mm (f) adapter 1250-1745 Opt BAB
2.4 mm (f) to Type N (f) adapter 11903B X
Overload Detector DAC Adjustment Setup