Specifications

50 MHz Calibrator Amplitude Adjustment
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
50 MHz Calibrator Amplitude Adjustment
An RF substitution method is made using the 50 MHz, -25 dBm calibrator Z5602A. The
level of the calibrator is read by the PSA and then the PSA’s internal 50 MHz calibrator is
measured. The difference is stored in memory as a correction.
Required Test Equipment
Test Equipment Model Number
E4440A
E4443A, E4445A
E4446A
E4448A
50 MHz Calibrator
Z5602A 0ption 51
(Opt H35 for BAB
Opt H24 for E4446A, E4448A)
X X
50 MHz Calibrator Amplitude Adjustment Setup