Specifications

Second LO Power Adjustment
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Second LO Power Adjustment
The 3.6 GHz 2
nd
LO Power is measured with a power meter at the A20 Lowband 2
nd
LO
Test Port, A20J6. The power is controlled by a DAC on the A13 Front End Driver. The
computer reads the power and adjusts the DAC for the proper level. The instrument cover
and top shield must be removed to access the LO Test Port, but it is not necessary to put
PC assemblies on extenders.
The instrument internally reads the power level with a detector and ADC. This is done to
keep the power level constant with changing temperature. The adjustment stores a value in
the FE Driver EEROM used to calibrate the ADC so that the value corresponds to the
proper power level. The 2
nd
LO Power needs to be adjusted from a conservatively low
level, and then gradually increased to the proper level in order to avoid saturating the 2
nd
LO amplifiers.
This adjustment should be performed after replacing the A20 and/or A13 assemblies.
Required Test Equipment