Specifications
PSA Test Equipment
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Test Equipment
Instrument Critical Specifications
Recommended
Agilent Model Number
Alternative Agilent Model
Number
Use
1
Signal Sources
Synthesized
Signal Generator
Frequency: 1 MHz to 2.5 GHz
Spectral purity:
7 MHz to 1321 MHz
Spurious (non-harmonic):
at 50 MHz ≤-90 dBc
≤ 2.56 GHz, < -78 dBc
Harmonics:
< -30 dBc ≤ +13 dBm
SSB Phase Noise:
≤ -106 dBc/Hz at 1 kHz offset
(or residual FM ≤ 0.5 Hz,
300 Hz to 3 kHz integration BW).
VSWR: < 1.5:1
8663A
A, P, T
Synthesized
Signal Generator
Frequency: 100 kHz to 6.0 GHz
Spectral Purity:
SSB Phase Noise @ 1 GHz,
20 kHz offset = -134 dBc/Hz
Harmonics:
-30 dBc @ ≤ +10 dBm output
8665B
(Option 004)
P
Sweep Signal
Generator
(2 required, only 1 needs to
be Option 008)
Frequency: 10 MHz to 26.5 GHz
Frequency Resolution: 1 Hz
Harmonic level: < -40 dBc
Amplitude range: -20dBm to +13
Amplitude resolution: 0.02
Harmonic level: < 2 GHz, < -30 dBc
≥ 2 GHz & < 26.5 GHz, < -50 dBc
VSWR: < 20 GHz, 1.6:1
≤ 31 GHz, 1.8:1
83630B
(Option 008)
83630A
(Option 008),
83640A/B
(Option 008),
83650A/B
(Option 008)
A, P, T
http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/testequip.htm (1 of 7) [5/29/2003 8:46:54 AM]