Specifications
Power Bandwidth Accuracy Performance Test
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Power Bandwidth Accuracy Performance Test
In this test, the power bandwidth function is checked at a CW frequency of 50 MHz for
resolution bandwidths from 1 Hz to 1.2 MHz in a 1, 3, 10 sequence. The signal source is
the internal 50 MHz calibrator signal. The power bandwidth function reads the integrated
power between a marker pair which enclose the CW signal.. This power is compared to
the maximum power of the CW signal as read by the E444XA marker.
Required Test Equipment
None
Troubleshooting
http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/powbw.htm [5/29/2003 8:47:09 AM]