Specifications

Residual Responses Performance Test
Agilent Technologies
PSA Series Spectrum Analyzers
Test and Adjustment Software
Residual Responses Performance Test
The PSA RF Input is terminated and the analyzer is swept from 200 kHz to 1 MHz. Then
the analyzer is swept in incremental 10 MHz spans from 1 MHz to the upper frequency
range. After each sweep, the Marker Peak function is used to measure the highest
amplitude. The amplitude and frequency of any response above the specification are
noted.
Required Test Equipment
Test Equipment Model Number
E4440A
E4443A, E4445A
E4446A
E4448A
50 ohm Termination 909A, Option 012 X
50 ohm Termination 85138B X
3.5 mm (f) to Type N (f) adapter 1250-1745 (Option BAB)
Residual Responses Test Setup
http://mktwww.soco.agilent.com/field/lp/PSA/PSAcalSWwebhelp/Performance/resid.htm (1 of 2) [5/29/2003 8:46:55 AM]