Technical data

What's Wrong?
1. Do you have any self test failures, running error
messages (beeping) or raw channel power
(unratioed power) problems? = MAGIC QUESTION.
2. Failed performance tests?
3. Display or Front Panel abnormalities, measurement
errors, or display hang-ups?
4. Low or no power symptoms?
No Obvious Failure Type
The best approach to repair the 8510XF at this point
is to verify that the 85101C is good. Once it is known
to be good, it will act as your diagnostic controller.
Remove the external controller, printer, etc. and the cables
from the 8510XF system.
Run the 8510XF Service Program to verify that the
8510XF is OK to a 95% confidence level. To do this,
run all the service program tests in the "85101C
Display/Processor Service Program menu (#1)."
These tests take about ten minutes. Refer to the
8510XF Service Program Flowchart. See partial
flowchart on next page.
NOTE: the above tests can be skipped if the 85101C
passes the power up 85101 self test. The self test
has only a 70% confidence level, so use it accordingly.
Do a Pre-Operational Check of the
8510XF System
Check Front and Rear Panel control settings, cables and
GPIB addresses. Check the 85101 free run switch
on the A5 Processor Board (all zeros), and proper firm-
ware revisions for source and 8510XF (you may need
to reload the Operating System in an extreme case).
Check that the 8510XF firmware is XF.01.00 or greater.
First, turn on the Sources, then the Test Set, then the
8510XF, or the system cabinet power switch. Then
check all troubleshooting LED's. See the 8510XF
system level troubleshooting block diagram for these
LED's. See installation for additional information in
performing the Preflight Checkout.
Read the 8510XF Calibration and Verification and
Cable manuals on how to inspect, gage, clean, and
use the calibration and verification devices, waveguide,
and test port return cables. Make sure, if your 85059A
Verification Kit has been certified to 110 GHz within the
past year.
Obvious Calibration or Verification Problem
Try the backup standard software by reloading the
Operating System firmware.
It might be a known software bug. See STARS bug report.
Call the factory with details for checkout and duplication
of unknown and new bugs!
Obvious Software Problem
CHECK POWER SUPPLIES
Check all main (110V/220V) fuses and power switches.
Remove 85101, 85102 and E7342A/52A instrument
covers and check LED power and service indicator lights.
Recheck by measuring each power supply voltage with a
DVM.
The 85102 power switch LED is supplied by the +5V
supply in the 85101.
Obvious Power Supply Problem
Obvious Self-Test, running error, or unratioed power test
failure. Refer to the 8510XF Service Manual, part
number E7350-90001, Service and Troubleshooting.
Refer to Qualified
Personnel Only!
83621A/B
LO Source
45 MHz to 20 GHz
83651A/B
RF Source
45 MHz to 50 GHz
X2
PREAMP
PREAMP
SLOPE AMP
SLOPE AMP
X3
50-75 GHz
50-75 GHz
50-75 GHz
2-50 GHz
2-50 GHz
RF
(Incident)
a1, a2
RF
(Reflected or
Transmitted)
test b1, b2
Test Port
( 1 of 2)
<50 GHz
<50 GHz
<50 GHz
<2 GHz
<2 GHz
Microwave Subassembly Millimeter Wave Subassembly
LO
LO
LO
LO
LO
8-18 dB
LO
t
t
r
3t
4t
4r
3r
2t
1t
1r
2r
LO
LO
75-110 GHz
45 MHz - 110 GHz
75-110 GHz
75-110 GHz
Combiner
Combiner
Splitter
Combiner
Splitter
r
Splitter
r
Splitter
Splitter
R
R
R
R
R
R
R
R
I
I
RF
RF
RF
RF
RF
RF
RF
45 MHz to
2 GHz
RF
IF
IF
IF
IF
IF
IF
IF, 75-110 GHz
Band
Option 005
50 GHz-
110 GHz
45 MHz-
50 GHz
IF, 75-110 GHz
Band
IF, 50-75 GHz
Band
IF, 50-75 GHz
Band
IF, 2-50 GHz
Band
IF, 2-50 GHz
Band
IF, 45 MHz-2 GHz
Band
Option 005
IF, 45 MHz-2 GHz
Band
Option 005
IF
IF
I
I
I
I
I
I
L
L
L
L
L
L
L
L
Reference IFsTest IFs
Test Head ( 1 of 2 )
Millimeter
Wave
Controller
Obvious Failure Type
Obvious Failure Type
Simplified Signal Path of Unratioed Power Level Tests
If NO Obvious Failure Type
Prove 85101 is OK
E7340
flowcht1.cdr
Rev. 2 4/22/98
8510XF System Service Flow Diagram (Sheet 1of 2)
To Sheet 2
EPROM ID on the
GPIB Board A12
5-17/5-18