Technical data
8510XF Service Quick Reference Guide
4-11
Performance Verification
Verification Procedures
9. Verify Cal Set for 45 MHz Measurement (If Required) is set to No.
10. Select
Done and follow the program prompts.
11. Select
Measure Data to measure the verification standards. Follow the
program prompts.
12. When verification has been completed view the data, then print or save
(if required). If needed the verification standard can be remeasured by
selecting:
Repeat Measure.
13. Select
Select Standard to return to the Standard Selection Menu. Repeat
steps 7 through 12 above for the other verification standard (refer to
Table 4-6 on page 4-9).
14. Select
Prior Menu until the Main Menu appears.
15. Select
Quit Program to exit the program, then close the program window to
finish.