Technical data

8510XF Service Quick Reference Guide
3-1
3 Making 8510XF Adjustments
NOTE This chapter covers both the CRT and LCD displays. Some display-related
adjustments apply only to instruments with CRT display.
The following table provides adjustments most frequently needed for
8510XF system service. Most of the 8510XF is self-adjusting except for the
following:
Table 3-1 Frequently used 8510XF adjustments
Procedure
1
1. Reference subsequent sections in this document.
Title Adjustment Function Assembly Adjusted
1 Vertical Alignment Adjustment
2
2. These adjustments apply to a CRT-based display only.
Aligns softkey labels and mechanical
softkey buttons.
A11
2 Degausing (Demagnetizing) the 85101
display
2
Demagnetize the display.
3 8360 Series Sources Full User CAL
4 Power Level Calibration Insures level power at the test ports Leveling DAC/ EEPROM, A16,
and Leveling Compensation
Board, A7