Technical data

85108L System Manual
System Service and Troubleshooting
Troubleshooting Strategy
6-4
8360 Series Source Check
Software, Revision, Language, and GPIB Address
When an 8360 series source is used with an 8510 based network analyzer system, it must be set to
the “analyzer” language.
The language and GPIB address of the 8360 series sources can be set using either of two methods:
rear panel switches, and front panel keys. The procedure given below sets the GPIB address to 19
(the address of RF source 1).
Rear Panel Switch Method
1. Turn on the source.
2. On the network analyzer, press:
3. Turn the source off.
SYSTEM MENU
[GPIB MENU]
[ANALYZER]
(an * will appear indicating the selection.)
[ADRS MENU] (top line of display shows the language and software revision: LANG:
ANALYZER, ADRS=nn, REV dd mm yy
[8360 ADRS] (top line of display shows: HP-IB ADDRESS:nn SELECTED)
Entry 1 9 ENTER (top line of display shows: HP-IB ADDRESS:19 SELECTED)