Table 6-1 System Setup Check Table
Check: Should be: Additional
information:
Rear panel settings Line voltage Set to local line
voltage
See rear panels of each
instrument
Fuse size Appropriate for the
line voltage
See rear panels of each
instrument
Operating system
firmware revision
8510C C.06.50 or higher See the label on the
operating system disk
85102B (if used) B.05.11 or higher See the label on the
operating system disk
8360 series 08 Mar.91 or later See 8360 series source
check
8510 system bus
address
RF source in 8360 LANG = 001
ADDRESS = 10010
(binary)
See 8360 series system
bus
LO source is 8360 LANG = 001
ADDRESS = 10010
(binary)
See 8340/41 source
check
RF source if 8340B
or 8341B
19 (decimal) address
check
See 8340/41 source
LO source if 8340B
or 8341B
18 (decimal) address
check
See 8340/41 source
Controller GPIB bus
address
3.5 inch drive 700 (decimal) See your disk drive
manual for GPIB
address switch location
8510 716 (decimal) Checked after power up