Technical data

85108L System Manual
System Installation
System Arrival
2-16
Firmware Revision
All properly functioning systems display the firmware revision information illustrated in Figure 2-8
after system turn on.
8510C-based systems must have firmware revision 8510C.06.50 or greater.
8510B-based systems must have firmware revision 8510B.05.11 or greater. If your 8510B has
firmware revision 8510B.06.00 or greater, and you are using an 836xx source, you must upgrade
the firmware in each 836xx source you are using (must be March 08, 1991, or higher).
Some 8510 systems may require later firmware revisions than those listed above. Please consult
your 8510C On-Site Service Manual and your Agilent representative for more information on
upgrading a network analyzer or source firmware (refer to “Contacting Agilent” on page iv).