Technical data
85108L System Manual
Loading the System Configuration Disk
C-2
The 85108L system configuration disk contains files for pulsed-RF configuration and two
nonpulsed-RF configurations. Refer to Table C-1 for 8510C firmware revision C.06.5x, or Table C-2
for 8510C firmware revision C.07.xx to select the appropriate file for your system.
Procedure
1. Insert the 85108L System Configuration disk into the disk drive of the network analyzer. On the
network analyzer, press
TAPE/DISC [LOAD] [MORE] [MACHINE DUMP]. Use the knob to select the
appropriate file, and press
[LOAD FILE].
2. Remove the disk when it has finished loading. Instrument State 8 (the power-up state) of the
loaded file will be recalled automatically.
Table C-1 85108L System Configuration Disk Files -
8510C Firmware Revision C.06.5x
1
to C.06.99 (p/n 85108-10007)
1. x denotes any digit as acceptable.
File Name Description of Configuration
Source(s) Detectors and mode
MD_PULS Multiple Wide BW pulsed-RF mode
MD_NPULS Multiple Normal BW nonpulsed-RF mode
MD_SINGSRC
2
2. This configuration requires a single source, and a test set that provides an LO signal such
as the 8514, 8515, 8516, or 8517. Refer to Chapter 9, “System Installation,” in the 8510C
On-Site Service Manual (p/n 08510-90282) for a configuration diagram.
Single Normal BW nonpulsed-RF mode
Table C-2 85108L System Configuration Disk Files -
8510C Firmware Revision C.07.xx
1
or higher (p/n 85108-10008)
1. x denotes any digit as acceptable.
File Name Description of Configuration
Source(s) Detectors and mode
MD_LOWPLS Multiple Wide BW pulsed-RF mode
MD_LOWNPLS Multiple Normal BW nonpulsed-RF mode
MD_SINGSRC
2
2. This configuration requires a single source, and a test set that provides an LO signal,
such as the 8514, 8515, 8516, or 8517.
Single Normal BW nonpulsed-RF mode