Technical data

85108L System Manual B-3
Hardware and Instrument States
Table B-3 Instrument States Specific to the Pulsed-RF System
System Parameter Pulsed-RF CW
Power source #1
+8 dBm
a
10 dBm
Power slope src 1 OFF OFF
Power source #2
10 dBm
b
N/A
Power slope src 2 2.5 dB / GHz N/A
Pulse config detector
wide BW
c
normal BW
a. Set source 1 power to the highest level possible without IF overload. Power must be set in
the “normal BW” mode, then changed to the “wide BW” mode, if desired.
b. Set source #2 power to 10 dBm if an 8349B amplifier is used to amplify the LO signal.
c. Set to normal BW for nonpulsed-RF operation.
Table B-4 Multiple Source Settings for the Pulsed-RF System
Define Settings
Source 1 Mult. Numer. = 1
Mult. Denom. = 1
Offset Frequency = 0.000000000 GHz
Source 2 Mult. Numer. = 1
Mult. Denom. = 1
Offset Frequency = 0.020000017 GHz
Receiver Constant frequency = 0.020000000 GHz