Technical data

4-6 8510C Service Quick Reference Guide
Performance Verification
CW Frequency Accuracy Test
NOTE Be sure to connect the test set output to the 500 MHz 26.5 GHz position
input on the frequency counter. Also set the input switch to the 500 MHz
26.5 GHz position.
6. Measure the frequency with the counter, and record the value on the test record as in
Table 4-2.
In Case of Difficulty
If the measured values do not meet the specifications listed on the test record, refer to your
source manual for adjustment and troubleshooting instructions.
Table 4-2 Performance Test Record for CW Frequency Accuracy Test
Instrument Model Report Number Date
Frequency Minimum
Specification
Recorded Results Maximum
Specification
Uncertainty
a
a. The measurement uncertainty is quoted for these performance tests using only the recommended
models specified in Table 8-1 of the On-Site Service Manual. The quoted uncertainty represents
limits of ± 3 times the equivalent standard deviation (3 σ) and is intended to represent a 90%
confidence level.
45 MHz
b
b. For all sources except the 83622A/B and 83624A/B.
44.999955 MHz 45.000045 MHz ± 10 Hz
2 GHz
c
c. For 83622A/B / 24A/B only.
1.999998 GHz 2.000002 GHz ± 10 Hz
20 GHz
d
d. For 83620A/B / 21A/B / 22A/B / 23A/B / 24A/B only.
19.99998 GHz 20.00002 GHz ± 4 kHz
26.5 GHz
e
e. For 83631A/B / 40A/B / 51A/B and 8340 only.
26.4999735 GHz 26.5000256 GHz ± 5 kHz
40 GHz
f,g
f. For 83640A/B only
g. For 83650A/B / 51A/B only.
39.999996 GHz 40.000004 GHz ± 5 kHz
50 GHz
g,h
h. 40 GHz and 50 GHz can be measured with a 26.5 GHz counter to verify 40 or 50 GHz due
to double function in source.
49.9999756 GHz 50.0000256 GHz ± 5 kHz