Technical data
2-4 8510C Service Quick Reference Guide
Instrument Compatibility
8360 Series Coupler to Bridge Detector
8360 Series Coupler to Bridge Detector
All 8510B/C systems when used with an 8360 series source that have a bridge detector
inside will potentially cause an over-modulation running error message. If your 8510C has
Rev.7.0 or greater, an RF unleveled caution message will also be displayed.
Workaround
Increase your source power and use test set attenuators, or use a source that has a coupler
in its leveling loop. The problem is caused by dc switching spikes from the test set RF pin
diode switch getting back into the sources leveling loop and causing an unleveled
condition. If this is not satisfactory, there are three upgrade kits available: the 8514B
/8515A /85110A (08510-60119), 8517B (08510-60118), and the 85110L (08510-60121) test
sets. The upgrade kit will eliminate the RF switching spikes from getting back into the
source leveling loop by installing a dc return between the RF input and RF pin diode
switch inside the test set. Contact an Agilent engineer for installation of these kits.
Agilent Model Coupler to Detector Conversion Prefix Determination
83620 Prefix equal to and greater than 3245A will cause problem
83622 Prefix equal to and greater than 3245A will cause problem
83630 Prefix equal to and greater than 3245A will cause problem
83650 Prefix equal to and greater than 3245A will cause problem
83621 Prefix equal to and greater than 3139A will cause problem
83631 Prefix equal to and greater than 3139A will cause problem
83651 Prefix equal to and greater than 3139A will cause problem
83623 Couplers only will not cause problem
83624 Couplers only will not cause problem